상세 보기
- Kim, Eunyoung;
- Shin, Beomsoo;
- Kwon, Kwang-il;
- Bang, Joon Seok;
- Kang, Wonku
WEB OF SCIENCE
1SCOPUS
1초록
We developed a method for the simultaneous quantification of 7-O-succinyl macrolactin A and its active metabolite, macrolactin A, in dog plasma. After protein precipitation with acetonitrile including flufenamic acid as an internal standard, 7-O-succinyl macrolactin A, macrolactin A, and flufenamic acid were chromatographed on a reverse-phase C-18 analytical column. The mobile phase, consisting of 20 mM acetate buffer and acetonitrile, was eluted using a gradient program at 1 mL/min, and the UV absorbance was measured at 230 nm. The retention times of 7-O-succinylmacrolactin A, flufenamic acid, and macrolactin A were 3.4, 4.8, and 6.9 min, respectively. The coefficient of variation in the assay precision for both substances was less than 6%, and the accuracy ranged from 96 to 105%. This method was used to measure the concentrations of 7-O-succinylmacrolactin A and macrolactin A in dog plasma following an intravenous administration of a single dose (25 mg/kg) of 7-O-succinyl macrolactin A salt.
키워드
- 제목
- Simultaneous determination of 7-O-succinyl macrolactin A and its active major metabolite, macrolactin A in dog plasma using high-performance liquid chromatography with UV detection
- 저자
- Kim, Eunyoung; Shin, Beomsoo; Kwon, Kwang-il; Bang, Joon Seok; Kang, Wonku
- 발행일
- 2014-10
- 유형
- Article
- 권
- 37
- 호
- 20
- 페이지
- 2833 ~ 2836