Selective Diagnosis of Diabetes Using Pt-Functionalized WO3 Hemitube Networks As a Sensing Layer of Acetone in Exhaled Breath

  • Choi, Seon-Jin; 
  • Lee, Inkun; 
  • Jang, Bong-Hoon; 
  • Youn, Doo-Young; 
  • Ryu, Won-Hee; 
  • 외 2명
Citations

WEB OF SCIENCE

292
Citations

SCOPUS

299

초록

Thin-walled WO3 hemitubes and catalytic Pt-functionalized WO3 hemitubes were synthesized via a polymeric fiber-templating route and used as exhaled breath sensing layers for potential diagnosis of halitosis and diabetes through the detection of H2S and CH3COCH3, respectively. Pt-functionalized WO3 hemitubes with wall thickness of 60 nm exhibited superior acetone sensitivity (R-air/R-gas = 4.11 at 2 ppm) with negligible H2S response, and pristine WO3 hemitubes showed a 4.90-fold sensitivity toward H2S with minimal acetone-sensing characteristics. The detection limit (R-air/R-gas) of the fabricated sensors with Pt-functionalized WO3 hemitubes was 1.31 for acetone of 120 ppb, and pristine WO3 hemitubes showed a gas response of 1.23 at 120 ppb of H2S. Long-term stability tests revealed that the remarkable selectivity has been maintained after aging for 7 months in air. The superior cross-sensitivity and response to H2S and acetone gas offer a potential platform for application in diabetes and halitosis diagnosis.

제목
Selective Diagnosis of Diabetes Using Pt-Functionalized WO3 Hemitube Networks As a Sensing Layer of Acetone in Exhaled Breath
저자
Choi, Seon-Jin; Lee, Inkun; Jang, Bong-Hoon; Youn, Doo-Young; Ryu, Won-Hee; Park, Chong Ook; Kim, Il-Doo
DOI
10.1021/ac303148a
발행일
2013-02
저널명
Analytical Chemistry
권
85
호
3
페이지
1792 ~ 1796