dc.description.abstract | As a standard approach to treating glaucoma, a silicone tube device is inserted into the eye to drain water and thereby lower the intraocular pressure (IOP). However, the fixed diameter of the tube often results in too much initial water drain. Suture-sheaving of the outer tube wall also leads to random IOP drops over time. In this study, self-expansion of the inner tube diameter was achieved by inserting a shape memory tube into the silicone tube. The difference in controlling small to large IOP drops before and after tube diameter expansion was demonstrated via computational modeling, a flow pumping system, and rabbit experiments. | - |
dc.format.extent | 7 | - |
dc.language | 영어 | - |
dc.language.iso | ENG | - |
dc.publisher | AMER CHEMICAL SOC | - |
dc.title | Shape Memory Tube Plug for Fine-control of Intraocular Pressure by Glaucoma Devices | - |
dc.type | Article | - |
dc.publisher.location | 미국 | - |
dc.identifier.doi | 10.1021/acsbiomaterials.0c00649 | - |
dc.identifier.scopusid | 2-s2.0-85090357426 | - |
dc.identifier.wosid | 000551355300006 | - |
dc.identifier.bibliographicCitation | ACS BIOMATERIALS SCIENCE & ENGINEERING, v.6, no.7, pp 3784 - 3790 | - |
dc.citation.title | ACS BIOMATERIALS SCIENCE & ENGINEERING | - |
dc.citation.volume | 6 | - |
dc.citation.number | 7 | - |
dc.citation.startPage | 3784 | - |
dc.citation.endPage | 3790 | - |
dc.type.docType | Article | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Materials Science | - |
dc.relation.journalWebOfScienceCategory | Materials Science, Biomaterials | - |
dc.subject.keywordPlus | AHMED GLAUCOMA | - |
dc.subject.keywordPlus | VALVE IMPLANT | - |
dc.subject.keywordPlus | DRAINAGE DEVICES | - |
dc.subject.keywordPlus | COMPLICATIONS | - |
dc.subject.keywordPlus | SURGERY | - |
dc.subject.keywordPlus | EYES | - |
dc.subject.keywordAuthor | glaucoma | - |
dc.subject.keywordAuthor | shape memory | - |
dc.subject.keywordAuthor | self-diameter expansion | - |
dc.subject.keywordAuthor | tube insertion | - |
dc.subject.keywordAuthor | intraocular pressure | - |
dc.identifier.url | https://pubs.acs.org/doi/10.1021/acsbiomaterials.0c00649 | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
Sookmyung Women's University. Cheongpa-ro 47-gil 100 (Cheongpa-dong 2ga), Yongsan-gu, Seoul, 04310, Korea02-710-9127
Copyright©Sookmyung Women's University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.