상세 보기
- Kim, Yeongjin;
- Na, Youngjin;
- Kim, Soobyeong;
- Yi, Jeongyoon;
- Ahn, Bummo
WEB OF SCIENCE
1SCOPUS
0초록
Stress incontinence occurs due to the inability to control an urge to urinate, which affects the quality of daily life. Although there is an existing therapeutics with polymer mesh tape, it is hard to fix around the urethra to restore the function of the pubourethral ligament. Therefore, in this research, an infra-obturator tension (IOT) sling was introduced to minimize the mobility of the bladder and urethra against intra-abdominal pressure. To verify the IOT performance, two different hypotheses were formulated: The hardness and volume of the IOT support the urethra and help prevent it from deforming by intra-abdominal pressure. The rotatory motion of the IOT can bend the urethra at the contact point and help restore the function of pubourethral ligament to increase vaginal tension. For the first hypothesis, a finite element simulation was conducted using three-dimensional geometrical model obtained by the computed tomography images of patients. For the second hypothesis, surgeons performed an IOT insertion operation and analyzed the sling rotation from the patients' CT images (mean: 21 degrees). From the results, the mobility of the urethra was decreased because of IOT insertion. The mobility was also decreased because of the subsequent fibrotic changes from the encapsulation of the IOT.
키워드
- 제목
- INFRA-OBTURATOR TENSION SLING FOR URINARY INCONTINENCE TREATMENT
- 저자
- Kim, Yeongjin; Na, Youngjin; Kim, Soobyeong; Yi, Jeongyoon; Ahn, Bummo
- 발행일
- 2019-05
- 유형
- Article
- 권
- 19
- 호
- 3